상세 보기
Commissioning and operation of 11.74T human MRI magnet for Gachon university gil medical center, South Korea
- Pittaluga, S.;
- Dormicchi, O.;
- Mauro, L.;
- Pizzigoni, G.;
- Ventura, D.;
- ... Chung, Jun-Young;
- ... Kim, Woo-Kyung;
- 외 1명
Citations
WEB OF SCIENCE
3Citations
SCOPUS
3초록
Magnetic resonance imaging (MRI) systems with magnetic field strengths higher than 7 Tesla (T), offer exceptional image resolution and signal-to-noise ratio, enabling more detailed visualization of fine anatomical structures, especially for brain, musculoskeletal, and cardiovascular studies. Only a few of them are operating in the whole world. An 11.74 T magnet for human MRI has been commissioned at Gachon University Gil Medical Center (GUGMC) in South Korea in 2022 and has now exceeded two-year operational life at full field. An overview of its features, the commissioning test campaign, and the experience gained after operations is presented.
키워드
superconducting magnets; UHF MRI; MRI
- 제목
- Commissioning and operation of 11.74T human MRI magnet for Gachon university gil medical center, South Korea
- 저자
- Pittaluga, S.; Dormicchi, O.; Mauro, L.; Pizzigoni, G.; Ventura, D.; Jo, Young Seung; Chung, Jun-Young; Kim, Woo-Kyung
- 발행일
- 2025-03
- 유형
- Article
- 권
- 38
- 호
- 3