상세 보기
Harmonic-Reduced Bias Circuit for Ultrasound Transducers
WEB OF SCIENCE
2SCOPUS
2초록
The gain of class-C power amplifiers is generally lower than that of class-A power amplifiers. Thus, higher-amplitude input voltage signals for class-C power amplifiers are required. However, high-amplitude input signals generate unwanted harmonic signals. Therefore, a novel bias circuit was proposed to suppress the harmonic signals generated by class-C power amplifiers, which improves the output voltage amplitudes. To verify the proposed idea, the input harmonic signals when using a harmonic-reduced bias circuit (−61.31 dB, −89.092 dB, −90.53 dB, and −90.32 dB) were measured and were found to be much lower than those when using the voltage divider bias circuit (−57.19 dB, −73.49 dB, −70.97 dB, and −73.61 dB) at 25 MHz, 50 MHz, 75 MHz, and 100 MHz, respectively. To further validate the proposed idea, the pulse-echo measurements were compared using the bias circuits. The peak-to-peak echo amplitude and bandwidth of the piezoelectric transducer, measured when using a harmonic-reduced bias circuit (27.07 mV and 37.19%), were higher than those achieved with a voltage divider circuit (18.55 mV and 22.71%). Therefore, the proposed scheme may be useful for ultrasound instruments with low sensitivity.
키워드
- 제목
- Harmonic-Reduced Bias Circuit for Ultrasound Transducers
- 저자
- Choi, Hojong
- 발행일
- 2023-04
- 유형
- Article
- 저널명
- Sensors
- 권
- 23
- 호
- 9
- 언어
- ENG
- 출판사
- MDPI
- 발행국가
- 스위스
- ISSN
- E 1424-8220