A Comparative Study of Birdcage RF Coil Configurations for Ultra-High Field Magnetic Resonance Imaging

Citations

WEB OF SCIENCE

10

초록

Improvements in transmission and reception sensitivities of radiofrequency (RF) coils used in ultra-high field (UHF) magnetic resonance imaging (MRI) are needed to reduce specific absorption rates (SAR) and RF power deposition, albeit without applying high-power RF. Here, we propose a method to simultaneously improve transmission efficiency and reception sensitivity of a band-pass birdcage RF coil (BP-BC RF coil) by combining a multi-channel wireless RF element (MCWE) with a high permittivity material (HPM) in a 7.0 T MRI. Electromagnetic field (EM-field) simulations, performed using two types of phantoms, viz., a cylindrical phantom filled with oil and a human head model, were used to compare the effects of MCWE and HPM on BP-BC RF coils. EM-fields were calculated using the finite difference time-domain (FDTD) method and analyzed using Matlab software. Next, to improve RF transmission efficiency, we compared two HPM structures, namely, a hollow cylinder shape HPM (hcHPM) and segmented cylinder shape HPM (scHPM). The scHPM and MCWE model comprised 16 elements (16-rad BP-BC RF coil) and this coil configuration demonstrated superior RF transmission efficiency and reception sensitivity along with an acceptable SAR. We expect wider clinical application of this combination in 7.0 T MRIs, which were recently approved by the United States Food and Drug Administration.

키워드

inductively coupled wireless element; high permittivity material; FDTD method; birdcage RF coil; 7; 0 T MRI; TO-NOISE RATIO; RADIOFREQUENCY POWER DEPOSITION; HIGH-PERMITTIVITY; ELECTROMAGNETIC-FIELDS; SURFACE COILS; SENSITIVITY; TRANSMIT; ARRAY; NMR; SIMULATIONS
제목
A Comparative Study of Birdcage RF Coil Configurations for Ultra-High Field Magnetic Resonance Imaging
저자
Seo, Jeung-Hoon; Han, Yeji; Chung, Jun-Young
DOI
10.3390/s22051741
발행일
2022-03
유형
Article
저널명
Sensors
권
22
호
5