Absorbed dose and image quality according to the scan mode and shield position in pediatric brain computed tomography scans

Citations

WEB OF SCIENCE

2
Citations

SCOPUS

2

초록

In this study, computed tomography dose index (CTDIvol), dose-length product (DLP), organ dose and signal-to-noise ratio (SNR) were measured in pediatric brain helical and volumetric CT scans using a pediatric phantom (equivalent to a 10-year-old) and optically stimulated luminescence dosemeters (OSLD) as the scan protocol used in Hospital A. Volumetric scans showed lower CTDIvol, DLP and organ dose than helical scans, and the SNR showed similar results. The organ doses were lower in the phantoms protected with front, rear or front and rear shielding than in those without shielding. However, no statistically significant differences were found among the different shield locations. In pediatric brain CT scans, to reduce exposure dose and patient discomfort while maintaining image quality, a volumetric scan rather than a helical scan was preferred, and a good shielding effect was observed with front or rear shielding.

키워드

CHILDRENCHEST
제목
Absorbed dose and image quality according to the scan mode and shield position in pediatric brain computed tomography scans
저자
Bae, EunmiCho, YoungkwonKim, Sungchul
DOI
10.1093/rpd/ncad001
발행일
2023-04
유형
Article
저널명
Radiation Protection Dosimetry
199
5
페이지
385 ~ 390