Fetal Bovine Serum을 포함한 세포 배양액에 담근 Indium Tin Oxide 전극 계면의 전기화학적 특성

Electrochemical Properties of Indium Tin Oxide Electrodes Immersed in a Cell Culture Medium with Fetal Bovine Serum

초록

For the biocompatibility test of implantable devices or for the sensitivity evaluation of biomedical sensors,it is required to understand the mechanism of the protein adsorption and the interaction between the adsorbed proteins and cells. In this study, the adsorption of proteins in a cell culture medium with fetal bovine serum onto an indium tin-oxide electrode was characterized by using linear sweep voltammetry and impedance spectroscopy. We immersed the fabricated ITO electrodes in the culture medium for 30, 60, or 90 min, and then measured the electrochemical properties of electrodes with 10 mM Fe(CN)63−/4− and 0.1 M KCl electrolyte. With an increase of contacting time, the anodic peak current was decreased and the charge transfer resistance was increased. However, both parameters were recovered to the values before contact with the medium after the treatment of Trypsin/Ethylenediaminetetraacetic acid hydrolyzing proteins.

키워드

Adsorption; Electrode interface; Impedance spectroscopy; Linear sweep voltammetry; Protein
제목
Fetal Bovine Serum을 포함한 세포 배양액에 담근 Indium Tin Oxide 전극 계면의 전기화학적 특성
제목 (타언어)
Electrochemical Properties of Indium Tin Oxide Electrodes Immersed in a Cell Culture Medium with Fetal Bovine Serum
저자
최원석; 조성보
DOI
10.9718/JBER.2013.34.1.34
발행일
2013-03
저널명
의공학회지
권
34
호
1
페이지