상세 보기
Optimization and Performance Evaluation of the Median Modified Wiener Filter Algorithm in Breast Electromagnetic X-ray Image
- Mo, Yunhee;
- Kang, Seong-Hyeon;
- Lee, Youngjin
WEB OF SCIENCE
1SCOPUS
1초록
The mask size of the median modified Wiener filter (MMWF) algorithm was optimized for application to digital mammographic phantom images obtained under different tube currents to obtain an optimal image by removing noise with a minimal dose of radiation. A simulation was conducted using the Female Adult Mesh (FASH) phantom to optimize the mask size of the MMWF algorithm. The optimal mask size was determined by measuring various evaluation factors and applying different mask sizes to the noisy image of the various evaluation phantom. An American College of Radiology (ACR) phantom image was obtained per milliamperes using digital mammography equipment to evaluate the MMWF algorithm with the optimal mask size. Based on the noise level and similarity evaluation factors, the optimal mask size for the MMWF algorithm was 7 x 7. The results of the experiment indicated that improvements in the noise level evaluation factors were most noticeable for the MMWF, followed by the median, Gaussian, and Wiener filters. Overall, we confirmed that the optimal mask size for the MMWF algorithm was 7 x 7 for application to digital mammographic images, and its effectiveness was proven through a comparative evaluation with conventional filters.
키워드
- 제목
- Optimization and Performance Evaluation of the Median Modified Wiener Filter Algorithm in Breast Electromagnetic X-ray Image
- 저자
- Mo, Yunhee; Kang, Seong-Hyeon; Lee, Youngjin
- 발행일
- 2023-12
- 유형
- Article
- 권
- 28
- 호
- 4
- 페이지
- 392 ~ 400
- 언어
- ENG
- 출판사
- KOREAN MAGNETICS SOC
- 발행국가
- 대한민국
- 분량
- 9 페이지
- ISSN
- E 2233-6656
P 1226-1750