Study of the Noise Reduction Algorithm with Median Modified Wiener Filter for T2-weighted Magnetic Resonance Brain Images

  • 최동혁; 
  • 강성현; 
  • 박찬록; 
  • 이영진
Citations

WEB OF SCIENCE

8
Citations

SCOPUS

9

초록

The purpose of this study was to confirm the utility of the median modified Weiner filter (MMWF) noise reduction algorithm in T2-weighted brain MR images. We acquired brain MR images using both real experiment and simulation, and performed comparative evaluation with conventional noise reduction algorithms through calculation of factors for the noise level and similarity. As a result, evaluation of the noise level and similarity showed the most improved in the image, which the MMWF noise reduction algorithm was applied. Moreover, additional experiment was conducted using real MR device and water phantom to more clearly prove the efficiency of the MMWF noise reduction algorithm. The noise level and intensity profile results derived from the real T2-weighted MR image proved the effectiveness of the MMWF noise reduction algorithm. In conclusion, the proposed MMWF noise reduction algorithm demonstrated a very promising performance for improving the image quality in T2-weighted MR image.

키워드

MRiLab simulation program; T2-weighted image; Median modified Wiener filter (MMWF); Noise reduction algorithm
제목
Study of the Noise Reduction Algorithm with Median Modified Wiener Filter for T2-weighted Magnetic Resonance Brain Images
저자
최동혁; 강성현; 박찬록; 이영진
DOI
10.4283/JMAG.2021.26.1.050
발행일
2021-03
저널명
Journal of Magnetics
권
26
호
1
페이지
50 ~ 59