Changing the entrance surface dose and image quality by applying an air mattress to the table commonly used in computed tomography

Citations

WEB OF SCIENCE

1
Citations

SCOPUS

1

초록

Background: In this study, the changes in the entrance surface dose (ESD) and image quality were examined when an air mattress was used to reduce the X-ray scattering caused by treatment tables commonly used for computed tomography (CT). Materials and Methods: Dual-energy CT (DECT) was used, with the ESD of Alderson Radiation Therapy (ART) phantom measured by scanning with different X-ray tube voltages and X -ray tube currents for three scenarios: when no air mattress was employed, and when a 5 cm or 10 cm thick air mattress was employed. The statistical significance of the changes in ESD and image quality were based on the presence and thickness of the air mattress. Additionally, the variations of the X-ray tube voltage and current for different air mattress thicknesses were investigated using paired t-test. Results: For all X-ray tube voltages and currents, applying an air mattress significantly improved both the ESD and image quality, with the 5 cm air mattress improving both the ESD and image quality. Conclusion: The 5 cm air mattress produced no artifacts in the diagnostic images, and demonstrated a statistically significant reduction in patient ESD during DECT imaging.

키워드

air mattress; computed tomography; SNR; Entrance surface dose; CNR; CARBON-FIBER; ATTENUATION
제목
Changing the entrance surface dose and image quality by applying an air mattress to the table commonly used in computed tomography
저자
Park, J. Y.; Kim, S.
DOI
10.52547/ijrr.20.2.25
발행일
2022-04
유형
Article
저널명
INTERNATIONAL JOURNAL OF RADIATION RESEARCH
권
20
호
2
페이지
425 ~ 429

파일 다운로드

Thumbnail