상세 보기
Integrated Low-Voltage Compliance and Wide-Dynamic Stimulator Design for Neural Implantable Devices
- Oh, Yeonji;
- Hong, Jonggi;
- Kim, Jungsuk
WEB OF SCIENCE
2SCOPUS
3초록
In this study, a pulse frequency modulation (PFM)-based stimulator is proposed for use in biomedical implantable devices. Conventionally, functional electrical stimulation (FES) techniques have been used to reinforce damaged nerves, such as retina tissue and brain tissue, by injecting a certain amount of charge into tissues. Although several design methods are present for implementing FES devices, an FES stimulator for retinal implants is difficult to realize because of the chip area, which needs to be inserted in a fovea, sized 5 mm x 5 mm, and power limitations to prevent the heat generation that causes tissue damage. In this work, we propose a novel stimulation structure to reduce the compliance voltage during stimulation, which can result in high-speed and low-voltage operation. A new stimulator that is composed of a modified high-speed PFM, a 4-bit counter, a serializer, a digital controller, and a current driver is designed and verified using a DB HiTek standard 0.18 mu m process. This proposed stimulator can generate a charge up to 130 nC, consumes an average power of 375 mu W during a stimulation period, and occupies a total area of 700 mu m x 68 mu m.
키워드
- 제목
- Integrated Low-Voltage Compliance and Wide-Dynamic Stimulator Design for Neural Implantable Devices
- 저자
- Oh, Yeonji; Hong, Jonggi; Kim, Jungsuk
- 발행일
- 2023-01
- 유형
- Article
- 저널명
- Sensors
- 권
- 23
- 호
- 1
- 언어
- ENG
- 출판사
- MDPI
- 발행국가
- 스위스
- ISSN
- E 1424-3210
P 1424-8220