Performance Evaluation of Image Characteristics by Changing Parameters in Brain T2 Weighted Neuroimages using MRiLab Simulation

Performance Evaluation of Image Characteristics by Changing Parameters in Brain T2 Weighted Neuroimages using MRiLab Simulation
  • 임준; 
  • 강성현; 
  • Chan Rok Park; 
  • 이영진
Citations

WEB OF SCIENCE

2
Citations

SCOPUS

2

초록

Research has been conducted on techniques for adjusting magnetic resonance imaging (MRI) parameters to acquire images with better characteristics. In this study, we aim to confirm the optimal parameter values through quantitative evaluation and tendency analysis of T2 weighted images of the human brain acquired by adjusting the flip angle, number of excitations (NEX), and bandwidth (BW) using the user-friendly MRiLab simulation program. We also considered noise level and similarity evaluation parameters. According to the results, as NEX increases and BW decreases for a flip angle of 90°, the noise level and similarity evaluation improved, whereas they became less optimal beyond a 90° flip angle. Excessive parameter variation was found to cause additional noise and artifacts in such images, resulting in a deterioration of image clarity. Therefore, we confirmed that setting appropriate parameters is critical in magnetic resonance image acquisition.

키워드

Magnetic resonance imaging (MRI); flip angle; Number of excitation (NEX) Bandwidth (BW); MRiLab simulation program; brain T2 weighted image
제목
Performance Evaluation of Image Characteristics by Changing Parameters in Brain T2 Weighted Neuroimages using MRiLab Simulation
제목 (타언어)
Performance Evaluation of Image Characteristics by Changing Parameters in Brain T2 Weighted Neuroimages using MRiLab Simulation
저자
임준; 강성현; Chan Rok Park; 이영진
DOI
10.4283/JMAG.2021.26.1.041
발행일
2021-03
저널명
Journal of Magnetics
권
26
호
1
페이지
41 ~ 49