상세 보기
Simulation Study of Radio Frequency Safety and the Optimal Size of a Single-Channel Surface Radio Frequency Coil for Mice at 9.4 T Magnetic Resonance Imaging
- Seo, Jeung-Hoon;
- Ryu, Yeunchul;
- Chung, Jun-Young
WEB OF SCIENCE
4SCOPUS
5초록
The optimized size of a single-channel surface radio frequency (RF) coil for mouse body images in a 9.4 T magnetic resonance imaging (MRI) system was determined via electromagnetic-field analysis of the signal depth according to the size of a single-channel coil. The single-channel surface RF coils used in electromagnetic field simulations were configured to operate in transmission/reception mode at a frequency of 9.4 T–400 MHz. Computational analysis using the finite-difference time-domain method was used to assess the single-channel surface RF coil by comparing single-channel surface RF coils of varying sizes in terms of |B1|-, |B1+|-, |B1−|- and |E|-field distribution. RF safety for the prevention of burn injuries to small animals was assessed using an analysis of the specific absorption rate. A single-channel surface RF coil with a 20 mm diameter provided optimal B1-field distribution and RF safety, thus confirming that single-channel surface RF coils with ≥25 mm diameter could not provide typical B1-field distribution. A single-channel surface RF coil with a 20 mm diameter for mouse body imaging at 9.4 T MRI was recommended to preserve the characteristics of single-channel surface RF coils, and ensured that RF signals were applied correctly to the target point within RF safety guidelines.
키워드
- 제목
- Simulation Study of Radio Frequency Safety and the Optimal Size of a Single-Channel Surface Radio Frequency Coil for Mice at 9.4 T Magnetic Resonance Imaging
- 저자
- Seo, Jeung-Hoon; Ryu, Yeunchul; Chung, Jun-Young
- 발행일
- 2022-06
- 유형
- Article
- 저널명
- Sensors
- 권
- 22
- 호
- 11
- 언어
- ENG
- 출판사
- MDPI
- 발행국가
- 스위스
- ISSN
- E 1424-8220
P 1424-8220