Pentafrequency Resonator based on a CRLH Microstrip for Multinuclear MRI and MRS at 7 T
Pentafrequency Resonator based on a CRLH Microstrip for Multinuclear MRI and MRS at 7 T
Citations
WEB OF SCIENCE
0Citations
SCOPUS
0초록
In this study, we present a transmission-line design based on the concept of composed right- and left-hand microstrip (CRLHTL) that is able to produce a uniform B1 field at five different frequencies. The design uses the same conductor line for all five frequencies, and maintains a uniform field along the Z axis. The CRLH concept is used to ensure that the line is operating with a uniform propagation constant in a lossless scenario. We performed electromagnetic simulations to validate the design.
키워드
pentafrequency resonator; magnetic resonance imaging/magnetic resonance spectroscopy; multinuclear; microstrip antenna
- 제목
- Pentafrequency Resonator based on a CRLH Microstrip for Multinuclear MRI and MRS at 7 T
- 제목 (타언어)
- Pentafrequency Resonator based on a CRLH Microstrip for Multinuclear MRI and MRS at 7 T
- 저자
- Daniel Hernandez
- 발행일
- 2020-12
- 권
- 25
- 호
- 4
- 페이지
- 547 ~ 555
- 언어
- ENG
- 출판사
- 한국자기학회
- 발행국가
- 대한민국
- 분량
- 9 페이지
- ISSN
- E 2233-6656
P 1226-1750